DEVELOPMENT OF PHASE STABILIZED SWEPT SOURCE OPTICAL COHERENCE TOMOGRAPHY FOR BIOMEDICAL IMAGING AND SENSING

dc.contributor.advisorLarin, Kirill V.
dc.contributor.committeeMemberSharma, Pradeep
dc.contributor.committeeMemberMetcalfe, Ralph W.
dc.contributor.committeeMemberFranchek, Matthew A.
dc.contributor.committeeMemberWood, Lowell T.
dc.creatorManapuram, Ravi 1981-
dc.date.accessioned2014-11-21T20:50:49Z
dc.date.available2014-11-21T20:50:49Z
dc.date.createdAugust 2012
dc.date.issued2012-08
dc.date.updated2014-11-21T20:50:50Z
dc.description.abstractSeveral life-threatening diseases could either be cured or be eased by diagnosing them during earlier stages. Currently, medical imaging is one of the most reliable methods of disease diagnostics. This dissertation describes the development of a medical imaging device that has a great potential for early diagnosis of various diseases. The developed system utilizes optical coherence tomography (OCT) and is capable of 3D imaging of tissues with near cellular resolution noninvasively. It is also capable of sensing minute changes in tissue refractive index or surface displacement via phase-sensitive measurements. The system showed an axial resolution of 8 µm, lateral resolution of 15 μm, an imaging depth of 9 mm, signal to noise ratio of 101 dB and a phase stability of 9 mrad. After development, the system was applied to several biomedical applications such as the detection of microbubbles in mice tails in vivo and mechanical wave propagation in mice corneas in vivo. In the live mice tails, microbubbles of sizes as small as 50 μm were detected, therefore indicating the capability of the system to serve as an early diagnostic tool for diseases caused by decompression sickness or gas emboli. Apart from imaging, this dissertation also describes a method to apply OCT for elastography applications. Results demonstrate that the system is capable of measuring amplitude of mechanical waves as small as 30 nm. The high sensitivity of the system was exploited to measure wave propagation in live mice corneas as a function of age. To the best of our knowledge, this is the first time OCT has been applied to measure wave parameters in ocular tissues in vivo.
dc.description.departmentMechanical Engineering, Department of
dc.format.digitalOriginborn digital
dc.format.mimetypeapplication/pdf
dc.identifier.urihttp://hdl.handle.net/10657/781
dc.language.isoeng
dc.rightsThe author of this work is the copyright owner. UH Libraries and the Texas Digital Library have their permission to store and provide access to this work. Further transmission, reproduction, or presentation of this work is prohibited except with permission of the author(s).
dc.subjectOptical coherence tomography (OCT)
dc.subjectPhase
dc.subjectPhD dissertation
dc.subjectOphthalmology
dc.subject.lcshMechanical engineering
dc.titleDEVELOPMENT OF PHASE STABILIZED SWEPT SOURCE OPTICAL COHERENCE TOMOGRAPHY FOR BIOMEDICAL IMAGING AND SENSING
dc.type.dcmiText
dc.type.genreThesis
thesis.degree.collegeCullen College of Engineering
thesis.degree.departmentMechanical Engineering, Department of
thesis.degree.disciplineMechanical Engineering
thesis.degree.grantorUniversity of Houston
thesis.degree.levelDoctoral
thesis.degree.nameDoctor of Philosophy

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
PhD_copy_08_15_2012_Final-1.pdf
Size:
6.58 MB
Format:
Adobe Portable Document Format

License bundle

Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
2.12 KB
Format:
Plain Text
Description: