Numerical and Experimental Study of Pacemaker Safety under MRI

dc.contributor.advisorChen, Ji
dc.contributor.committeeMemberJackson, David R.
dc.contributor.committeeMemberWilton, Donald R.
dc.contributor.committeeMemberZouridakis, George
dc.contributor.committeeMemberKainz, Wolfgang
dc.creatorFeng, Shi
dc.date.accessioned2018-03-05T20:29:47Z
dc.date.available2018-03-05T20:29:47Z
dc.date.createdDecember 2014
dc.date.issued2014-12
dc.date.submittedDecember 2014
dc.date.updated2018-03-05T20:29:48Z
dc.description.abstractPacemaker safety issues with MRI (Magnetic Resonance Imaging) are discussed in this thesis. In order to analyze the electromagnetic compatibility issue of pacemaker/defibrillator with MRI RF (radio frequency) coil and human body model, novel techniques for the evaluation of MRI-induced heating or induced voltage in the vicinity of implanted metallic lead have been developed based on the reciprocity theorem. The measurement configurations have been described in detail. The technique was validated in both numerical and experimental studies on different lead structures. More than one hundred transfer functions of real practical pacemakers are measured, and hundreds of validation tests were performed on those transfer functions. In addition, the effects of human body/phantom were also discussed based on the measured transfer functions and human body simulations. This approach allows one to decouple the micro-scale metallic lead simulation/measurement from the macro-level phantom human simulations within the MRI scanners. Consequently, the estimation of MRI-induced heating on an implanted lead, and the induced voltage on the pacemaker device can be greatly simplified. The analysis about the pacemaker safety issue will become very efficient. In addition, this method clearly explains the induced lead heating mechanism during MRI procedures.
dc.description.departmentElectrical and Computer Engineering, Department of
dc.format.digitalOriginborn digital
dc.format.mimetypeapplication/pdf
dc.identifier.urihttp://hdl.handle.net/10657/2806
dc.language.isoeng
dc.rightsThe author of this work is the copyright owner. UH Libraries and the Texas Digital Library have their permission to store and provide access to this work. Further transmission, reproduction, or presentation of this work is prohibited except with permission of the author(s).
dc.subjectMRI
dc.subjectPacemaker
dc.subjectEMC
dc.subjectSynthetic aperture radar (SAR)
dc.subjectInduced voltage
dc.subjectFDTD
dc.subjectReciprocity
dc.titleNumerical and Experimental Study of Pacemaker Safety under MRI
dc.type.dcmiText
dc.type.genreThesis
thesis.degree.collegeCullen College of Engineering
thesis.degree.departmentElectrical and Computer Engineering, Department of
thesis.degree.disciplineElectrical Engineering
thesis.degree.grantorUniversity of Houston
thesis.degree.levelDoctoral
thesis.degree.nameDoctor of Philosophy

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
FENG-DISSERTATION-2014.pdf
Size:
36.61 MB
Format:
Adobe Portable Document Format

License bundle

Now showing 1 - 1 of 1
No Thumbnail Available
Name:
LICENSE.txt
Size:
1.81 KB
Format:
Plain Text
Description: